This Technically Advance Balancing Machine is ideal for weighing both in Carat and Gold equally. It is well suited for your Precious Jewelry Items or Gemstones or metals such as Gold. Max Capacity 1,500g/Accuracy (Readability) 0.01g
Ohaus Scale SJX1502/E 1500/0.01gm is a Portable Balance ideal for Jewelry Weighing. It is equipped with lot of features that makes it durable and long lasting and not prone to overload or any other short comings of other normal weighing or Balancing Machines. Some of its few features are - Fast Stabilization Time, Multiple Weighing Units, Superior Overload Protection mechanism and many more.
Maximum Capacity | 1,500 g |
Readability | 0.01 g |
Pan Size | 5.5 in x 6.7 in (140 mm x 170 mm) |
Specifications : Ohaus Scale SJX1502/E 1500/0.01gm
Maximum Capacity | 1,500 g |
Readability | 0.01 g |
Pan Size | 5.5 in x 6.7 in (140 mm x 170 mm) |
Battery Life | 80 Hours with Disposable Batteries |
Communication | RS232 (Available as an Accessory); USB Host (Available as an Accessory); USB Device (Available as an Accessory); Ethernet (Available as an Accessory) |
Auxiliary Display Model | Available as an Accessory |
Dimensions | 2.1 in x 8.8 in x 8 in (54 mm x 224 mm x 202 mm) (HxLxW) |
Display | Large backlit LCD |
Internal Calibration | Not Applicable |
In-use cover | Available as an Accessory |
Linearity ± | 0.03 g |
Net Weight | 2.2 lb (1 kg) |
Pan Construction | Stainless Steel |
Power | AC Adapter (Included); 4 AA (LR6) Batteries (Not Included) |
Repeatability (typical) | 0.02 g |
Stabilization Time | 1.5 s |
Tare Range | To capacity by subtraction |
Transportation Case | Available as an Accessory |
Units of Measurement | Carat; Grain; Gram; Hong Kong Tael; Kilogram; Ounce; Ounce Troy; Pennyweight; Pound; Singapore Tael; Taiwan Tael; Tical; Tola |
Working Environment | 50°F – 104°F, 80%RH, non-condensing (10°C – 40°C, 80%RH, non-condensing) |
Dimension and weight | ||||
---|---|---|---|---|
Length | Width | Height | weight (Kgs) | |
Product | 22.4 | 20.2 | 5.4 | 1.00 |
Package | 30 | 26 | 9 | 1.40 |
Calibration Operation :
When the balance is operated for the first time, a span calibration is recommended to ensure accurate weighing results. Before performing the calibration, be sure to have the appropriate calibration weights. Ensure that the Security switch is set to unlocked position.
Press and hold Menu until [mMeNU] (Menu) is displayed. When the button is released, the display will show [C.A.L]. Press Yes to accept. [SpaN] will then be shown. Press Yes to begin the span calibration.
[--C--] will be displayed while zero reading is stored. Next, the display shows the calibration weight value. Press the No key toggle the value. Place the specified calibration mass on the pan. [--C--] will be displayed while the reading is stored. The display will show [done] if the calibration was successful. The balance returns to the previous application mode and is ready for use
Initial Calibration :
When the Balance is first installed, and when it is moved to another location, it must be calibrated to ensure accurate weighing results.
For SJX/E models, the balances must be calibrated by using the external calibration weights.
For SJX models, the balances have built in internal calibration system which can calibrate the balance without the need for external calibration masses. Alternately, the SJX balances can be manually calibrated with external masses. Have the appropriate calibration masses available before beginning calibration.